Sažetak

Background/Aim: This study aimed to evaluate the precision of the Raypex 6 apex locator in locating the apical constriction (AC) and major foramen (MF) during a root canal treatment compared with a microscopic evaluation. Another aim of the present study was to evaluate the accuracy of the Raypex 6 in the presence of different irrigating solutions (NaOCl, saline, EDTA, etidronic acid-HEBP). Material and Methods: One hundred and nine patients were included in this study and were randomly assigned into four groups according to the irrigation solution used; NaOCl, EDTA, HEBF and saline. Electronic MF (EMF) and AC (EAC) were measured by using Raypex 6. The teeth were extracted. The apical 3 mm of each canal was trimmed to expose the file tip. The samples were observed under a stereomicroscope, and the actual length of MF (ALMF) and AC (ALAC) were measured. The data were analyzed by using x2 test, and significance was set at P< 0.05. Results: The Raypex 6 was accurate 71.4% of the time to ±0.5mm and 93.3% of the time to ±1mm in determining the ALAC. While similar ALAC-EAC differences were observed in EDTA, NaOCl, and saline groups (p= 0.230), the highest differences were seen in the HEBP group (p= 1.000). The precision of Raypex 6 in determining the working length measurement depends on the type of irrigation. All solutions allowed reliable detection of AC. However, HEBF significantly increased the risk of overpreparation. Conclusions: Raypex 6 can be recommended for clinical use and its accuracy is not affected by the type of irrigant when locating MF.

raypex 6electronic apex locatorsworking length measurementirrigation solutions
Izvor i prava

Objavljeno u Balkan Journal of Dental Medicine pod licencom CC BY. Autorska prava zadržavaju autori.

Zapis časopisa u DOAJ-u (ISSN 2738-0807), provereno 28.09.2026.

Verzija od zapisa kod izdavača
Kako citirati: Peker Begum Unsal, Hepsenoglu Yelda Erdem, Ersahan Seyda, Eyuboglu Tan Firat. Accuracy of working length measurement by Raypex 6: Electronic apex locator versus actual measurements under stereomicroscope. Balkan Journal of Dental Medicine. 2022;26(1):15-21. doi:10.5937/bjdm2201015P

Autori

Peker Begum Unsal

Istanbul Medipol University, Faculty of Dentistry, Department of Endodontics, Istanbul, Turkey

Hepsenoglu Yelda Erdem

Istanbul Medipol University, Faculty of Dentistry, Department of Endodontics, Istanbul, Turkey

Ersahan Seyda

Istanbul Medipol University, Faculty of Dentistry, Department of Endodontics, Istanbul, Turkey

Eyuboglu Tan Firat

Istanbul Medipol University, Faculty of Dentistry, Department of Endodontics, Istanbul, Turkey