The effects of bar holder material, cantilever and vertical misfit on stresses in implant supported overdentures: Three dimensional finite element analysis
Kaymak Sümeyra · Yerliyurt Kaan
Rad u časopisuBalkan Journal of Dental Medicine, 26(2), 2022, str. 93–99
Sažetak
Background/Aim: The aim of this study is to examine with finite element analysis the distal bar extension, the bar substructure material type and the amount of bar substructure-abutment mismatch, and the stress caused by the implant at the surrounding bone tissue in bar-retained prostheses. Material and Methods: A bar-retained prosthesis model has been designed on three implants placed in the fully toothless lower jaw at the places of both canines and the midline. Bar holder according to distal cantilever lengths was modeled to be 0 mm, 8 mm and 14 mm. The vertical incompatibility of the bar holder substructure with the abutment was modeled to be 0 µm, 100 µm and 200 µm. A total of twenty-seven (3x3x3) different models were obtained with three different bar infrastructure materials (titanium, gold and chromium-cobalt). 150 N occlusal force was applied to the central fossa of the left 1st molar tooth with a rigid food stuff. Results: In the cortical bone, the highest maximum principle stress value (2.78 MPa) was analyzed around the anterior implant socket in the model 13 (gold, cantilever 0mm, misfit 100 µm). The highest von Mises stress value (343.43 MPa, which occurred at the selected joints in bar holders) was observed in model 27 (chrome-cobalt, cantilever 14mm, misfit 200 µm). Conclusions: When the length of the cantilever is 14 mm, it causes a significant increase in stress around the implant, especially near the cantilever. It has been observed that bar infrastructures with high elastic modulus create higher stress values.
Objavljeno u Balkan Journal of Dental Medicine pod licencom CC BY. Autorska prava zadržavaju autori.
Zapis časopisa u DOAJ-u (ISSN 2738-0807), provereno 28.09.2026.
Verzija od zapisa kod izdavačaAutori
Ankara, Turkey
Tokat Gaziosmanpaşa University, Department of Prosthodontics, Faculty of Dentistry, Tokat, Turkey
Srodni radovi
Po zajedničkim ključnim rečima.
How Do Differences of Dental Implants’ Internal Connection Systems Affect Stress Distribution? A 3-Dimensional Finite Element Analysis (2019) Are informations of subperiosteal implant procedure on YouTube really useful? (2025) Evaluation of beam hardening artifacts around dental implants: CT study on bovine ribs (2023) Detection of the effects of the various denture cleansing methods on overdenture retentive attachments with a scanning electron microscope (2022) Evaluation of Stress Distribution on all-on-four Concept and Conventional Implant Designs: 3D Finite Element Analysis (2021) Esthetic Outcome and Oral Health-Related Quality of Life after Restoration with Single Anterior Maxillary Implants (2021)